Polarization Efficiency Analysis: November 2006 run - part I

For the November run I tried measuring the polarization efficiency and HWP zero angle with a single calibration grid in 3 separate set-ups:

  1. M3 → aperture → perpendicular grid → SHARP → SHARC-II
  2. M3 → perpendicular grid → aperture → SHARP → SHARC-II
  3. M3 → grid ~22.5° from perpendicular → aperture → SHARP → SHARC-II
The details of my data analysis for this data can be found in my document on the July 2006 polarization efficiency measurements.

The plots below show an interesting systematic in this measurement. The data on H-V and S = (H-V)/(H+V) vs. HWP angle show a measurable offset along the y-axis. In addition, the H+V vs. HWP angle shows a distinct periodicity. My working theory had been that this is due to differences in the H and V gains (the January 2006 RGM has been applied to the data). That is, the polarization f-factor is clearly not unity. However, this hypothesis does consistently agree with all the data. Or rather, the offset varies dramatically from dataset to dataset, which is not what I would have expected for the f-factor. I have outlined the theory below anyway in the hopes that someone else might have some more ideas.

Consider the orthogonal signals are given by:
H = 0.5 I Gh [ P (cos 4θ + 1) + (1-P) ] and
V = 0.5 I Gv [ P (1 - cos 4θ) + (1-P) ]
where I is the total unpolarized intensity, P is the polarization, θ is the HWP angle, and Gh,v are the H and V array gains. I have assumed the arbitrary phase is zero for simplicity.

Now we have
H - V = 0.5 I [ P (Gh + Gv) cos 4θ + (Gh - Gv) ],
H + V = 0.5 I [ P (Gh - Gv) cos 4θ + (Gh + Gv) ]

From these equations we can see that if Gh = Gv then H(θ) - V(θ) and S(θ) should be symmetric about zero and H(θ) + V(θ) should be a constant. The data are clearly not consistent with the assumption Gh = Gv (especially for the low polarization found for set-up 2). The H(θ) - V(θ) and S(θ) data are consistent with the relations above. However, the H(θ) + V(θ) data exhibit a periodicity which differs from the expected 90°, but it is also clearly not random. I currently have no explanation for this.

The zero-offsets and H+V period are consistent between the two tests taken while the calibration grid wires were near-vertical. However, when the wires are horizontal, not only does the polarization level change, but so does the odd periodicity of H+V.

The H+V periodicity is not as obvious as in the differential data, but I think it is there. The differential data is much noisier than the non-differential data; I think the periodicity is just buried in the noise. I strongly suspect that much of this noise is due to the cold load not being replaced in the same spot (distance from detectors) for each HWP angle. I suggest blocking the cold load with warm eccosorb rather than removing/replacing the LN2 dewar for such differential tests in the future.

If the gain-difference hypothesis alone is correct I would have expected to get similar measurements for the f-factor despite the different measurement techniques. It may be that there is just a large error on the offset terms. But right now I can't rule out something else.


Figures above --- Left: Data for set-up #2 with calibration grid wires near-horizontal. Right: Data for set-up #3 with calibration grid wires near-vertical. Note that I have left the error bars off this plot and out of the fit so the parameter uncertainties and χ2 reported in the plots are not realistic.


Last updated by John Vaillancourt. 2006-Nov-21.
Return to SHARP analysis logbook.
Return to SHARP home page.