For the July 2008 observing run, as for previous runs, we measured the HWP zero angle with one wire grid using a differential measurement with a cold and warm load. We used the 350 μm HWP and filter for these tests. The optical path was
The results, shown in the figure below, yield P = 83 ± 5 %, δ = 6.2° ± 0.2°, and an offset of ~ 2%. The uncertainties reported here (but not in the figure) are estimated by fitting two subsets of the data, each with half the data points and comparing them to the fits with all data points.
.
The polarization efficiency of P = 83 ± 5% measured here is consistent with the meausurements from 2 wire grids (e.g., November and July 2006) since 1.00 - (1.00 - 0.83)^2 = 97%.
Using a value of δ=6.2°=96.2° (measured from the vertical) from the above data we find
V_null_encoder = 96.2°.
V_null_rel = 96.2 - 50 = 46.2°
HWP zero angle = 2 × 46.2 = 92.4°