For the November 2007 observing run, as for previous runs, we measured the HWP zero angle with one wire grid using a differential measurement with a cold and warm load. We used the 350 μm HWP and filter for these tests. The optical path was
The fit results, shown in the figure below, yield P = 89 ± 1 %, δ = 5.5° = 95.5° ± 0.2°, and an offset of ~ 5%. The uncertainties reported here (but not in the figure) are estimated by fitting two subsets of the data, each with half the data points.
.
Using a value of δ=5.5°=95.5° (measured from the vertical) from the above data we find
V_null_encoder = 95.5°.
V_null_rel = 95.5 - 50 = 45.5°
HWP zero angle = 2 × 45.5 = 91°