Characteristic | Value | Units | Notes | |
---|---|---|---|---|
Detector Operating Temperature | 285 | mK | measured at Northwestern | |
Beam FWHM | predicted: 4.1 measured: < 6.0 |
arcmin | predicted as follows: add (in quadrature) the pixel spacing (3.5 arcmin) and the diffraction-limited beam FWHM of approx. (0.45 mm/0.7 m) | |
Detector Noise | 50 | nV/Hz1/2 | measured with spectrum analyzer at 3.1 Hz | |
Responsivity (S) | 4 K blankoff wheel closed | 5.0*108 | V/W | varies by about 20% from detector to detector |
blankoff wheel open 300 K load |
Austral Winter 2000: 3.8*108
after 2001 upgrade: 2.1*108 | V/W | varies by about 20% from detector to detector | |
blankoff wheel open 200 K load (theoretical; interpolated) |
Austral Winter 2000: 4.1*108
after 2001 upgrade: 2.9*108 |
V/W | Interpolated using result of empirical fit to pre-2001 vs. 2001
load curves
that gives (Sclosed - Sopen) = (const.)(load power)0.83 |
|
Noise Equivalent Power | (atmospheric) | Austral Winter 2000: 1.10*10-16[1-e-tau*sec(z)]
1/2
after 2001 upgrade: 1.88*10-16[1-e-tau*sec(z)] 1/2 |
W/Hz1/2 | Fundamental noise from a 200K atmosphere with optical depth tau at 450 microns and zenith angle z |
(detector) | 1.2*10-16 | W/Hz1/2 | detector NEP = (det. noise)/S | |
Noise Equivalent Flux Density | (atmospheric) | Austral Winter 2000: 62*[e2*tau*sec(z)-etau*sec(z)
]1/2
after 2001 upgrade: 27*[e2*tau*sec(z)-etau*sec(z)]1/2 |
Jy/Hz1/2 | theoretical estimates based on 4' beam |
(detector) | Austral Winter 2000: 68*etau*sec(z)
after 2001 upgrade: 17.4*etau*sec(z) |
Jy/Hz1/2 | theoretical estimates based on 4' beam | |
Instrumental Transmission (cold optics) | Austral Winter 2000: 1.87
after 2001 upgrade: 5.5 |
% | determined from load curves with BOW closed vs. BOW open (Note this is measured with unpolarized light, so a perfect polarimeter would have 50% transmission in each of the two corresponding orthogonally-polarized detectors.) | |
Overall Instrumental Transmission | Austral Winter 2000: 1.12
after 2001 upgrade: 4.4 |
% | Cold Optics transmission times warm optics transmission (pressure window [80%]; black polyethylene OVCS filter [75%] that was used in 2000 but not 2001). Again, 50% is theoretical maximum transmission. | |
Fractional Bandpass | 0.1 | (none) | Bandpass centered at 450 microns |