SHARP Set-up Notes for January 2009. John Vaillancourt and Hua-bai Li -------------------------------------- HWP rotation and read-outs are confirmed to be working. SHARC-II alignments - Files 045821 (exit aperture) and 045822 (entrance aperture). - sharcslice was used to make images of the array. - fitgauss was used to find the centroids below: - exit center: X = 16.23, Y = 6.53 (off-center by 0.3, 0.0 pixels) - entrance center: X = 16.89, Y = 6.48 (off-center by 0.4, 0.0 pixels) - no chnages made to SHARC-II alignment SHARP alignment: - File 045825 = entrance aperture - File 045826 = exit aperture - Centroid offsets from nominal center for exit from fitgauss: H = (0.77 L, 0.22 U) V = (0.58 L, 0.34 U) HV mis-alignment at exit: 0.19 LR, 0.12 UD pixels - Centroid offsets from nominal center for entrance from fitgauss: H = (0.34 L, 0.19 U) V = (0.64 L, 0.03 U) HV mis-alignment at entrance: 0.30 LR, 0.16 UD pixels - Parallelism: H = 0.22 LR, 0.02 UD pixels V = 0.03 LR, 0.16 UD pixels