Setting up the Cryocontroller
Introduction to the Equipment
The following is a description of the equipment necessary to run the
cryomonitor for SPARO.
- PC- The computer which runs the software and saves the data collected
from SPARO's monitoring systems.
- Computer Breakout Box- this box provides connections directly into
the Labview Cards in the PC.
- Thermometry Multiplexing Box(TMB)- This box basically controls which
of the ROX 4-lead thermometers is being read by the lock-in based upon
a two bit TTL signal from the PC.
- Breakout Box #1(BB1)- This box essentially provides a patch bay for the
signals coming from the Cycling Cable from SPARO.
- Breakout Box #2(BB2)- This box handles all of the miscellaneous cables
run through Viper's cable wrap into the control room. It provides connections
from these cables to easy to handle BNC connectors.
- Lock-in Amplifier- This handles the reading of the resistances of the
ROX Thermometers using 4-lead techniques.
- Tan HP Power supplies- One of these supplies is used to heat the charcoal
pumps. The other is used to provide 24 V excitation to the He-3 pressure
sensors.
- Power supply for AD590's- This small green power supply is attached to
the AD590 budbox and is used to provide the 10 V excitation to these
temperature devices.
-
Blue Power Supplies- These are used to power the heaters in the Preamps.
- 10 microamp current source- This 9V battery powered 4-channel device is
used to power the Silicon Diode Thermometers on SPARO's charcoal pumps, OVCS,
and JFET boxes.
Ruthenium Oxide Thermometers
Specification sheets
The BNC connectors from Breakout Box#1(BB1) marked Detector(Inner) Stage V and I
should be connected to the Thermometry Multiplexing Box(TMB) at the connectors
marked V4 and I4, respectivey. Likewise The Pumped Pot V and I on BB1 should
go to the TMB connectors marked V3 and I3, respectively.
The TMB connections are as follows: Vout should be connected to the
input A of the lock-in amplifier. Iout gets connected to the Sine out on the
Lock-in through the 199.9k resistor box. TTL 1 and TTL 2 of the TMB get connected
to the correspondingly labeled BNC connectors on the Computer Breakout Box and
the 5V BNC on the TMB gets its power from the 5V BNC from the Computer Breakout
Box.
The lock-in amplifier measures the resistance of the ROX devices using a four
lead resistance measurment. The voltage the lock-in outputs to the cryocontroller
PC can be converted into a resistance via the following formula:
RT=[VLabViewRL/G Vsine out]/[1- VLabView/(GVsine out)]
G is the gain the lock-in gives to the output signal. It is defined as 10/sensitivity
.
The standard SPARO settings for the lock-in dictate the following:
Sensitivity: 5mV
Vsine out: 0.040V
Time Constant: 300ms
Coupling: AC
Input: A
ground: GROUND
Reserve: Normal
Filters: Both on
RL: 199.9 k
Frequency: 935.7 Hz
Auto-phase: on
Display: R, theta
Using this setup, the above formula reduces to:
RT=[2500 VLabView]/[1-(VLabView/80)]
Note that the denominator is a small correction which takes into account that
the current is limited not just by the load resistor, but also by the resistance
of the device itself.
Silicon Diode Thermometers
Specification sheets
The basic setup for these thermometers is to put 10 microamps through them
and measure a voltage. So for both of these thermometers, you need the
battery powered 10 microamp current source(it contains 4 BNC connectors and
4 switches, and sits on the shelf in the middle of the SPARO rack).
Place a BNC "T" on the current source and connect one end to BB1 at the
Charcoal Pumps Silicon Diode connection. Run the other end to the Computer
Breakout Box (to Channel 4). Hook up a similar connecton using one of the
other channels of the current source from the OVCS diode output of BB1 to channel
6 of the Computer Breakout Box.
Charcoal Pump Heaters
Using the top HP power supply located in the middle of the rack on the shelf,
run 3 BNC's in parallel. Connect one each to the two heater inputs on BB1
(Guard{outer} and Detector{inner} heaters). Run the other connection to
the Computer Breakout Box, Channel 7.
3He Pressure Sensors
Power each of the two 24 V inputs of Breakout Box #2 using parallel connections
from the lower HP power supply. Then, connect the signal BNC's to the Computer
Breakout Box in channel 11(for the inner stage) and channel 12(for the outer
stage).
Set the HP power supply to 24.0 V.
If you have to use the laboratory pressure sensor cables (i.e. if you are doing a cycle
in the lab), the color convention is that the red wire is the 24 V excitation,
the green wire gives the signal, and the white wire is the common.
Pumped Pot Pressure
In the Viper hallway, connect the Capacitance manometer readout to the
capacitance manometer cable via the DB 9 connector. Then run the DC out
output into the control room via a BNC. Connect this directly to the Computer
Breakout Box using Channel 10.
Preamp heat
Connect the BNC's on BB2 marked preamp heat(1 and 2) to the blue power supplies
located in the bottom of the rack. Note that the heat added is proportional to
the square of the voltage.
Preamp Gain control
Connect TTL 3 on the Computer Breakout Box to A1 on BB2. Then connect TTL 4 on
the Computer Breakout Box to A0 on BB1. The preamp gains which correspond to the
TTL settings are listed below.
Gain |
A1 |
A0 |
100 |
hi |
hi |
1000 |
hi |
lo |
10,000 |
lo |
hi |
100,000 |
lo |
lo |
Pump Resistor Thermometer and HWP Encoder
Connect the BNC connector from BB1 marked Pump Resistor Thermometer to an ohmmeter.
Likewise, connect the output marked encoder signal to another ohmmeter.
Reading the AD590's on the Pump and Preamps
Specification sheets
The AD590 budbox is taped to a power supply. Power the budbox as marked using
10 Volts from the power supply. Connect an ammeter to the BNC on the top of the budbox(set to microamps) and connect the final BNC to the breakout connection from the AD590. For the
preamps, these connections are located on BB2. For the Pump, there is a free BNC
connection labeled "Pump AD590". The ammeter reading in microamps is the
Kelvin temperature of the device.
Last updated: February 4,
1999 DTC