- The parameters used for the initialization are stored in the following file:
~viper/sparo2003/indexer/indexer.tcl
-
In that file, you will find the following lines:
tellindexer "I 100"
tellindexer "V 150"
-
To slow the motor to a speed of, say 70 steps per second, you should replace the
values 100 and 150 with the value 70, so that these lines will read as follows:
tellindexer "I 70"
tellindexer "V 70"
-
The minimum values for I and V are 20 steps per second.
-
In addition, sometimes stepper motors have a tendency to resonate more at lower speeds. To
prevent this, the IMC controller uses "variable resolution". You should make sure that
the indexer is set in variable resolution mode by adding the following line to the
initialization code:
tellindexer "H 1"
-
You can add this line right after you set the speeds, so the complete list of
initialization commands
will look like this:
tellindexer " "
tellindexer "K 10 10"
tellindexer "I 70"
tellindexer "V 70"
tellindexer "H 1"
tellindexer "Y 0 8"
tellindexer "O"
tellindexer "D 0"